NA100
Piston Nebulizer
convenient & effective
convenient & effective
Product Features
Particle size created by our VAT bottle is around 2.19μ m(MMAD) tested by Cascade Impactor and 3.221μm tested at Dv(50) by Malvern Spraytec. Compared to most bottles in the market, our proprietary bottle ensures more efficient and effective performance.
- MMAD ≤ 2.2 μm; Fine Particle Dose (FPD): 80-85%
- Consistent fine particle size for efficient respiratory treatment
- Patented Valve Adjustable nebulizer bottle (VAT)
- Powerful Piston compressor
- Mouthpiece and masks for adult and child included
- Built-in Compartment for Accessories
-
How to use the NA100?
-
Valve Adjustable Technology
Controllable nebulization rate at consistent particle sizeThe proprietary adjustable valve is able to deliver medications of different viscosity level according to every user’s conditions without exchanging parts. The Valve Adjustable Technology (VAT) bottle allows users to adjust different levels of Nebulization rate 0.15 (closed) / 0.4 (fully opened) ml/min at consistent particle size. Higher nebulization rate (fully opened) is for higher viscosity medications and higher breathing capacity user while lower nebulization rate with closed valve will be more appropriate for kids/infants with lower breathing capacity. VAT is recommended for respiratory disorders such as Chronic Obstructive Pulmonary Disease (COPD) and asthma as well as intensive use in general.Aerosol Performance Tested with 0.9% saline Cascade Impactor Malvern Spraytec MMAD FPD Dv(50) FPD 2.19μm 80.75% 3.221μm 77.38%